ABOUT US
 
 
WALK-IN-HUMIDITY CHAMBERS
 
WALK-IN-HUMIDITY CHAMBERS
 
Temperature Range : 20°C to 60°C Accuracy : ± 0.2°C
Humidity Range : 40% RH to 95% RH Accuracy : ±2% RH
Temperature Uniformity : ±1°C Humidity Uniformity : ± 3% RH

  • Temperature & Humidity Suitable for 25 °C& 60% RH, 40 °C & 75% RH, 30 °C & 65% RH

  • Specially designed chambers to carry out stability studies at low humidity conditions such as 25°C & 40% RH, 40°C & 25% RH NMT, 30°C & 35% RH
CONSTRUCTION :

The unit is a double walled chamber, with glass wool / PUF insulation. Interior is made of Stainless Steeland Exterior is either made of Mild Steel finished in powder coating paint (powder coating on both sides of the mild steel) or Stainless Steel. The unit has full-length inner glass door and outside metal door with magnetic gasket and lock. The unit is provided with perforated Stainless Steel trays. Forcedair circulation for uniform temperature. PU wheels for easy move ability.

INTELLIGENT CONTROL SYSTEM :

The controlling system having is PLC Base and PC communication on Ethernet and data acquisition system complying with 21 CFR Part 11

HEATING SYSTEM :

Long lasting stainless steel tubular heaters with s.s. fins.

HUMIDITY SYSTEM :

Humidity is by steam injection. Humidity system consist of boiler tank, which is fitted with two nos. of boiler heaters, one heater works as a stand by and automatic switch over to stand by heater if regular fails and give alarm.A water reservoir tank is connected to the boiler tank. DM water is preferred. A magnetic float switch provided for safety for boiler low level to cut off the its supply.

SENESOR :

Temperature and Humidity transmitter 4-20mA used for measurement. Temperature in degree Celsius and Humidity in %RH with 0.1°C and 0.1%RH resolution.

COOLING SYSTEM :

CFC free, the cooling system consists of hermetically sealed compressor coupled with evaporation coil and condenser, safe guarded by time delay circuit. Refrigerant 134A.

ALARMS :

Audio and visual alarms are generated for temperature / humidity variation, utility failure.

SAFETY :

The unit is provided with safety devices for temperature, and humidity over shoot in case of malfunction.

STANDBY SENSOR :

When controlling sensor fails it automaticallyswitch over to standby sensor.


The unit works on 230 Volts AC and 50Hz mains supply.
 
SELECTION GUIDE :

Models
Capacity
Inner Chamber
Dimension
External
Dimension
Trays
Exterior M.S.
Exterior S.S.
Cuft.
Liters
H x W x D (Mtr)
H x W x D+ 0.8 (Mtr)
Nos.
NEC 2190R
NEC 2190RS
190
5,400
2.00 x 1.36 x 2.00
2.12 x 1.48 x 2.12
24
NEC 2230R
NEC 2230RS
230
6,400
2.40 x 1.36 x 2.00
2.52 x 1.48 x 2.12
32
NEC 2280R
NEC 2280RS
280
8,000
2.00 x 2.00 x 2.00
2.12 x 2.12 x 2.12
24
NEC 2340R
NEC 2340RS
340
9,600
2.40 x 2.00 x 2.00
2.52 x 2.12 x 2.12
32
NEC 2530R
NEC 2530RS
530
15,000
2.40 x 2.50 x 2.50
2.52 x 2.62 x 2.62
48
NEC 2640R
NEC 2640RS
640
18,000
2.40 x 2.50 x 3.00
2.52 x 2.62 x 3.12
48

* N.B. Custom build chambers also available

OPTIONALS :

  • Stand by Cooling System (Auto Switch Over). The whole cooling system can be provided as standby. The stand by system comes in operational automatically with alarm when problem with regular cooling system is detected.

  • Common Software

  • Temperature and Humidity Scanner

  • Mobile Alert System

 

 

ETHERNET COMMUNICATION PLC BASED INTELLIGENT CONTROL SYSTEM
    ................................................
  

STANDARD HARDWARE AVAILABLE GLOBALLY
    ................................................
 
  

REMOTE MONITORING
 ................................................
  

 ALARMS & REMOTE SMS
 ................................................

   more features - functions - flexibilty



 
  Site Developed By DreamSoft | Home | About Us | Products | Features | Software | Support | Careers | Awards | Contact Us | Link